Purpose: Inadvertent stapling of intraluminal devices is a recognized but under-reported hazard, described mainly in bariatric surgery. Loss of tactile feedback in robotic surgery may delay its recognition. We demonstrate recognition, rescue and prevention of this event during robotic gastrectomy and propose a classification.
Methods: Two patients undergoing robotic gastrectomy for gastric cancer are presented. Case 1: an 82-year-old man undergoing proximal gastrectomy, in whom an oesophageal temperature probe was stapled at the gastro-oesophageal junction. Case 2: a 53-year-old woman undergoing distal gastrectomy, in whom the nasogastric tube was stapled during duodenal transection.
Results: Both events were recognised intraoperatively. In case 1, the staple line was opened, the probe retrieved intact, and the planned stapled oesophagogastric anastomosis replaced by a handsewn anastomosis. In case 2, the staple line was opened, the tube tip retrieved and the duodenal stump closed; the tube was withdrawn before further firing. Both patients recovered uneventfully. We propose a four-tier classification of accidental intraluminal tube stapling, based
on location of the accidental stapling. A four-step pre-firing check, clues for recognition of accidental stapling, and a six-step rescue sequence are also described.
Conclusions: Stapled intraluminal tubes are preventable. A spoken check with the anaesthesia team before every firing, early recognition and a structured rescue allow safe salvage without added morbidity.